Author: Lyoo, Kwang-Soo; Yeom, Minjoo; Kim, Jungho; Kim, Donghyuk; Ha, Gunwoo; Na, Woonsung; Le, Van Phan; Song, Daesub
Title: Development of rapid immunochromatographic strip test for the detection of porcine epidemic diarrhoea virus Document date: 2017_12_2
ID: 4szmu1dh_18
Snippet: Of the 130 swab samples tested, 58 were positive for PEDV and 72 were negative according to the ICA kit, whereas 60 were positive and 70 were negative according to real-time RT-PCR (Table 3) . Thus, three samples were positive (threshold cycle value of 29.5, 30.2 and 31.7) by real-time RT-PCR, but not detected in the ICA kit test. The relative sensitivity and specificity of the ICA kit compared with those of real-time RT-PCR were 95.0 per cent (9.....
Document: Of the 130 swab samples tested, 58 were positive for PEDV and 72 were negative according to the ICA kit, whereas 60 were positive and 70 were negative according to real-time RT-PCR (Table 3) . Thus, three samples were positive (threshold cycle value of 29.5, 30.2 and 31.7) by real-time RT-PCR, but not detected in the ICA kit test. The relative sensitivity and specificity of the ICA kit compared with those of real-time RT-PCR were 95.0 per cent (95 per cent confidence interval: 86.08-98.96) and 98.6 per cent (95 per cent confidence interval: 92.30-99.96), respectively.
Search related documents:
Co phrase search for related documents- confidence interval and cycle value: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22
- confidence interval and positive threshold cycle value: 1, 2, 3, 4
- confidence interval and real time: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- confidence interval and relative sensitivity: 1, 2, 3, 4, 5, 6
- confidence interval and specificity relative sensitivity: 1, 2
- confidence interval and swab sample: 1, 2, 3, 4, 5, 6, 7
- confidence interval and threshold cycle value: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21
- cycle value and positive threshold cycle value: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- cycle value and real time: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- cycle value and relative sensitivity: 1
- cycle value and specificity relative sensitivity: 1
- cycle value and swab sample: 1, 2, 3, 4, 5
- cycle value and test swab sample: 1
- cycle value and threshold cycle value: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- ICA kit and real time: 1, 2, 3, 4, 5, 6, 7, 8
- ICA kit and relative sensitivity: 1, 2, 3
- ICA kit and specificity relative sensitivity: 1, 2, 3
- ICA kit specificity relative sensitivity and relative sensitivity: 1, 2
- ICA kit specificity relative sensitivity and specificity relative sensitivity: 1, 2
Co phrase search for related documents, hyperlinks ordered by date