Author: Stohr, Joep J.J.M.; Wennekes, Meike; van der Ent, Martijn; Diederen, Bram M.W.; Kluytmans-van den Bergh, Marjolein F.Q.; Bergmans, Anneke M.C.; Kluytmans, Jan A.J.W.; Pas, Suzan D.
Title: Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens Cord-id: 03jc4kzc Document date: 2020_11_8
ID: 03jc4kzc
Snippet: INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer. METHODS: Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharynge
Document: INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer. METHODS: Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharyngeal samples in either E-swab® or GLY- were tested using the SARS-CoV-2 assay on the cobas®6800 System and compared to a lab developed assay. Additonally, SARS-CoV-2 RNA stability was assessed after one freeze-thaw cycle with or without lysis buffer. RESULTS: In total, 221 (58.3%) oro- and nasopharyngeal swabs, 131 (34.6%) deep respiratory specimens, and n = 25 (6.6%) swabs of unknown origin were included to study clinical performance. Only 4 samples gave discrepant results, all being positive in the LDA and not the cobas®6800 system. For stability testing, 66 samples without and 110 with lysis buffer were included. No clinically significant difference was found in test results after one freeze-thaw cycle and addition of lysis buffer. CONCLUSION: Based on our findings, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results as the LDA in oro-/nasopharyngeal swabs and deep respiratory specimens. Moreover, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results before and after a freeze-thaw cycle, with better preservation of low viral loads in lysis buffer.
Search related documents:
Co phrase search for related documents- acid extraction and acute respiratory syndrome coronavirus: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- acute respiratory syndrome coronavirus and additional sample: 1, 2, 3, 4, 5, 6, 7, 8
- acute respiratory syndrome coronavirus and low respiratory: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- acute respiratory syndrome coronavirus type and low respiratory: 1, 2
Co phrase search for related documents, hyperlinks ordered by date