Author: Arnaout, Ramy; Lee, Rose A.; Lee, Ghee Rye; Callahan, Cody; Yen, Christina F.; Smith, Kenneth P.; Arora, Rohit; Kirby, James E.
Title: SARS-CoV2 Testing: The Limit of Detection Matters
  • Cord-id: 3hazdvbl
  • Document date: 2020_6_4
  • ID: 3hazdvbl
    Snippet: Resolving the COVID-19 pandemic requires diagnostic testing to determine which individuals are infected and which are not. The current gold standard is to perform RT-PCR on nasopharyngeal samples. Best-in-class assays demonstrate a limit of detection (LoD) of ~100 copies of viral RNA per milliliter of transport media. However, LoDs of currently approved assays vary over 10,000-fold. Assays with higher LoDs will miss more infected patients, resulting in more false negatives. However, the false-ne
    Document: Resolving the COVID-19 pandemic requires diagnostic testing to determine which individuals are infected and which are not. The current gold standard is to perform RT-PCR on nasopharyngeal samples. Best-in-class assays demonstrate a limit of detection (LoD) of ~100 copies of viral RNA per milliliter of transport media. However, LoDs of currently approved assays vary over 10,000-fold. Assays with higher LoDs will miss more infected patients, resulting in more false negatives. However, the false-negative rate for a given LoD remains unknown. Here we address this question using over 27,500 test results for patients from across our healthcare network tested using the Abbott RealTime SARS-CoV-2 EUA. These results suggest that each 10-fold increase in LoD is expected to increase the false negative rate by 13%, missing an additional one in eight infected patients. The highest LoDs on the market will miss a majority of infected patients, with false negative rates as high as 70%. These results suggest that choice of assay has meaningful clinical and epidemiological consequences. The limit of detection matters.

    Search related documents:
    Co phrase search for related documents
    • lod detection and low concentration: 1, 2, 3, 4, 5, 6