Author: Andrew J Robinson; William London; Lazslo Kotan; Warwick Downing
Title: A modification to the Maquet Flow-i anaesthesia machine for ICU ventilation Document date: 2020_4_11
ID: l3d1dc93_33
Snippet: Leak tests. The successful prototype passed System Check when installed on six different Flow-i machines in Lakes DHB's Theatre complex. The average leak was 42ml/min, well within the <150ml/min allowed. Each machine was then restored to stock and immediately passed System Check normally before being returned to use......
Document: Leak tests. The successful prototype passed System Check when installed on six different Flow-i machines in Lakes DHB's Theatre complex. The average leak was 42ml/min, well within the <150ml/min allowed. Each machine was then restored to stock and immediately passed System Check normally before being returned to use.
Search related documents:
Co phrase search for related documents, hyperlinks ordered by date