Author: Andrew J Robinson; William London; Lazslo Kotan; Warwick Downing
Title: A modification to the Maquet Flow-i anaesthesia machine for ICU ventilation Document date: 2020_4_11
ID: l3d1dc93_18
Snippet: Total Loss with active humidification too wet. When our normal ICU ventilators are used with F&P active gas heating and humidification, we do not suffer significant rainout in the patient circuit. The Flow-i was therefore set up with an appropriate circuit and allowed to operate as a total loss system over night. This was done once without scavenging (AGS), and on the second night with AGS. In the AGS condition, the breathing system remained clea.....
Document: Total Loss with active humidification too wet. When our normal ICU ventilators are used with F&P active gas heating and humidification, we do not suffer significant rainout in the patient circuit. The Flow-i was therefore set up with an appropriate circuit and allowed to operate as a total loss system over night. This was done once without scavenging (AGS), and on the second night with AGS. In the AGS condition, the breathing system remained clear of liquid water, but condensation formed throughout the Volume Reflector. Without AGS, it was possible to decant liquid water from the Volume Reflector after a mere 12 hours.