Author: Xiong, Yujuan; Li, Zhen-Zhen; Zhuang, Qi-Zhen; Chao, Yan; Li, Fei; Ge, Yi-Yuan; Wang, Yi; Ke, Pei-Feng; Huang, Xian-Zhang
Title: Comparative performance of four nucleic acid amplification tests for SARS-CoV-2 virus Cord-id: 8m6793vd Document date: 2020_3_29
ID: 8m6793vd
Snippet: Coronavirus disease 2019 (COVID-19) can be screened and diagnosed through the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) by real-time reverse transcription polymerase chain reaction. SARS-CoV-2 nucleic acid amplification tests (NAATs) have been rapidly developed and quickly applied to clinical testing during the pandemic. However, studies evaluating the performance of these NAAT assays are limited. We evaluated the performance of four NAATs, which were marked by th
Document: Coronavirus disease 2019 (COVID-19) can be screened and diagnosed through the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) by real-time reverse transcription polymerase chain reaction. SARS-CoV-2 nucleic acid amplification tests (NAATs) have been rapidly developed and quickly applied to clinical testing during the pandemic. However, studies evaluating the performance of these NAAT assays are limited. We evaluated the performance of four NAATs, which were marked by the Conformité Européenne and widely used in China during the pandemic. Results showed that the analytical sensitivity of the four assays was significantly lower than that claimed by the NAAT manufacturers. The limit of detection (LOD) of Daan, Sansure, and Hybribio NAATs was 3000 copies/mL, whereas the LOD of Bioperfectus NAATs was 4000 copies/mL. The results of the consistency test using 46 samples showed that Daan, Sansure, and Hybribio NAATs could detect the samples with a specificity of 100% (30/30) and a sensitivity of 100% (16 /16), whereas Bioperfectus NAAT detected the samples with a specificity of 100% (30/30) and a sensitivity 81.25% (13/16). The sensitivity of Bioperfectus NAAT was lower than that of the three other NAATs; this finding was consistent with the result that Bioperfectus NAAT had a higher LOD than the three other kinds of NAATs. The four above mentioned reagents presented high specificity; however, for the detection of the samples with low virus concentration, Bioperfectus reagent had the risk of missing detection. Therefore, the LOD should be considered in the selection of SARS-CoV-2 NAATs.
Search related documents:
Co phrase search for related documents- accuracy evaluate and acid amplification test: 1, 2
- accuracy evaluate and acute respiratory syndrome: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19
- accuracy evaluate and acute respiratory syndrome coronavirus: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17
- accuracy evaluate and lod detection: 1
- acid amplification and actual lod: 1
- acid amplification and acute respiratory syndrome: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- acid amplification and acute respiratory syndrome coronavirus: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- acid amplification and lod detection: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18
- acid amplification test and acute respiratory syndrome: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- acid amplification test and acute respiratory syndrome coronavirus: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24
- acid amplification test and lod detection: 1
- acid extraction system and actual lod: 1
- acid extraction system and acute respiratory syndrome: 1, 2, 3
- acid extraction system and acute respiratory syndrome coronavirus: 1, 2, 3
- acid extraction system and lod detection: 1, 2, 3, 4
- actual lod and lod detection: 1, 2
- acute respiratory syndrome and lod detection: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
- acute respiratory syndrome and low virus concentration: 1, 2, 3
- acute respiratory syndrome coronavirus and lod detection: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
Co phrase search for related documents, hyperlinks ordered by date