Title: Saliva Is a Valid Alternative to Nasopharyngeal Swab in Chemiluminescence-Based Assay for Detection of SARS-CoV-2 Antigen Cord-id: hpl298y2 Document date: 2021_4_2
ID: hpl298y2
Snippet: Diagnostic methods based on SARS-CoV-2 antigens detection are a promising alternative to SARS-CoV-2 RNA amplification. We evaluated the automated chemiluminescence-based Lumipulse(®) G SARS-CoV-2 Ag assay on saliva samples, using Simplexa™ COVID-19 Direct assay as a reference test. Analytical performance was established on a pool of healthy donors’ saliva samples spiked with the 2019-nCoV/Italy-INMI1 isolate, whereas clinical performance was assessed on fresh saliva specimens collected from
Document: Diagnostic methods based on SARS-CoV-2 antigens detection are a promising alternative to SARS-CoV-2 RNA amplification. We evaluated the automated chemiluminescence-based Lumipulse(®) G SARS-CoV-2 Ag assay on saliva samples, using Simplexa™ COVID-19 Direct assay as a reference test. Analytical performance was established on a pool of healthy donors’ saliva samples spiked with the 2019-nCoV/Italy-INMI1 isolate, whereas clinical performance was assessed on fresh saliva specimens collected from hospitalized patients with suspect or confirmed COVID-19 diagnosis. The limit of detection (LOD) was 0.65 Log TCID50/mL, corresponding to 18,197 copies/mL of SARS-CoV-2 RNA. Antigen concentrations and SARS-CoV-2 RNA were highly correlated (r = 0.99; p < 0.0001). Substantial agreement (80.3%) and significant correlation (r = −0.675; p = 0.0006) were observed between Lumipulse(®) G assay results and Ct values on clinical samples, with 52.4% sensitivity and specificity 94.1%. Sensitivity exceeded 90.0% when calculated on samples with Ct < 25, and specificity was 100% when excluding samples from recovered patients with previous COVID-19 diagnosis. Overall, chemiluminescence-based antigen assay may be reliably applied to saliva samples to identify individuals with high viral loads, more likely to transmit the virus. However, the low positive predictive value in a context of low SARS-CoV-2 prevalence underscores the need for confirmatory testing in SARS-CoV-2 antigen-positive cases.
Search related documents:
Co phrase search for related documents- lod detection and low prevalence: 1, 2
- lod detection and low sensitivity: 1, 2, 3, 4, 5, 6
- long persistence and low prevalence: 1
- low prevalence and lumipulse sars: 1
- low prevalence and lumipulse sars ag: 1
Co phrase search for related documents, hyperlinks ordered by date