Author: Babiker, Ahmed; Ingersoll, Jessica M.; Adelman, Max W.; Webster, Andrew S.; Broder, Kari J.; Stittleburg, Victoria; Waggoner, Jesse J.; Kraft, Colleen S.; Woodworth, Michael H.
Title: Validation of High-Sensitivity Severe Acute Respiratory Syndrome Coronavirus 2 Testing for Stool—Toward the New Normal for Fecal Microbiota Transplantation
  • Cord-id: fxoi3aoi
  • Document date: 2021_6_9
  • ID: fxoi3aoi
    Snippet: INTRODUCTION: Mounting evidence demonstrates potential for fecal–oral transmission of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). The US Food and Drug Administration now requires SARS-CoV-2 testing of potential feces donors before the use of stool manufactured for fecal microbiota transplantation. We sought to develop and validate a high-sensitivity SARS-CoV-2 reverse transcriptase polymerase chain reaction (RT-PCR) procedure for testing stool specimens. METHODS: A modified e
    Document: INTRODUCTION: Mounting evidence demonstrates potential for fecal–oral transmission of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). The US Food and Drug Administration now requires SARS-CoV-2 testing of potential feces donors before the use of stool manufactured for fecal microbiota transplantation. We sought to develop and validate a high-sensitivity SARS-CoV-2 reverse transcriptase polymerase chain reaction (RT-PCR) procedure for testing stool specimens. METHODS: A modified extraction method was used with an RT-PCR assay adapted from the Centers for Disease Control and Prevention PCR protocol for respiratory specimens. Contrived specimens were created using pre-COVID-19 banked stool specimens and spiking in known concentrations of SARS-CoV-2-specific nucleic acid. The highest transcript concentration at which 2/2 or 1/2 SARS-CoV-2 targets were detected in 9/10 replicates was defined as the dual-target limit and single-target limit of detection, respectively. The clinical performance of the assay was evaluated with stool samples collected from 17 nasopharyngeal swab RT-PCR-positive patients and 14 nasopharyngeal RT-PCR-negative patients. RESULTS: The dual-target and single-target limit of detection were 56 copies/μL and 3 copies/μL, respectively. SARS-CoV-2 was detected at concentrations as low as 0.6 copies/μL. Clinical stool samples from known COVID-19-positive patients demonstrated the detection of SARS-CoV-2 in stool up to 29 days from symptom onset with a high agreement with nasopharyngeal swab tests (kappa statistic of 0.95, P value < 0.001). DISCUSSION: The described RT-PCR test is a sensitive and flexible approach for the detection of SARS-CoV-2 in stool specimens. We propose an integrated screening approach that incorporates this stool test to support continuation of fecal microbiota transplantation programs.

    Search related documents:
    Co phrase search for related documents
    • active replication and low concentration: 1
    • lod detection limit and low concentration: 1, 2, 3, 4, 5, 6
    • lod detection limit and lysis buffer: 1, 2, 3
    • long period and low concentration: 1, 2
    • long period and low proportion: 1
    • low concentration and lysis buffer: 1, 2
    • lysis buffer and ma waltham: 1, 2, 3
    • lysis buffer and ma waltham thermo fisher: 1, 2