Title: Transcutaneous carbon dioxide monitoring accurately predicts arterial carbon dioxide partial pressure in patients undergoing prolonged laparoscopic surgery.
  • Cord-id: b4r8lazz
  • Document date: 2010_1_1
  • ID: b4r8lazz
    Snippet: BACKGROUND There may be large differences between measurements of end-tidal carbon dioxide partial pressure (Petco(2)) and arterial carbon dioxide partial pressure (Paco(2)) during laparoscopic surgeries. Transcutaneous carbon dioxide (Ptcco(2)) monitoring can be used to noninvasively and continuously estimate Paco(2). In the present study we evaluated the accuracy of Ptcco(2) monitoring in predicting the Paco(2) during laparoscopic surgeries with prolonged pneumoperitoneum. METHODS Sixteen pati
    Document: BACKGROUND There may be large differences between measurements of end-tidal carbon dioxide partial pressure (Petco(2)) and arterial carbon dioxide partial pressure (Paco(2)) during laparoscopic surgeries. Transcutaneous carbon dioxide (Ptcco(2)) monitoring can be used to noninvasively and continuously estimate Paco(2). In the present study we evaluated the accuracy of Ptcco(2) monitoring in predicting the Paco(2) during laparoscopic surgeries with prolonged pneumoperitoneum. METHODS Sixteen patients who underwent laparoscopic radical gastrectomy or radical proctectomy under general anesthesia were included in the study. Their Paco(2), Petco(2), and Ptcco(2) values were measured at 3 time points before and after pneumoperitoneum. Agreement among measures was assessed by the Bland-Altman method. RESULTS Forty-eight sample sets were obtained. The average Paco(2)- Ptcco(2) difference was -0.9 + or - 6.4 mm Hg (mean + or - 2 SD). The average Paco(2) - Petco(2) difference was 7.5 + or - 7.0 mm Hg (mean + or - 2 SD). Paco(2) - Ptcco(2) was less than or equal to + or -5 mm Hg for 88% of the samples. Paco(2) - Petco(2) was less than or equal to + or -5 mm Hg for 17% of the samples (P < 0.05). CONCLUSIONS While undergoing long-term pneumoperitoneum laparoscopic surgery, Ptcco(2) monitoring is more accurate than is PETCO(2) monitoring in predicting the patients' Paco(2).

    Search related documents:
    Co phrase search for related documents
    • Try single phrases listed below for: 1