Title: Evaluation of absorbable and permanent mesh fixation devices: adhesion formation and mechanical strength.
  • Cord-id: oma1al3o
  • Document date: 2011_1_1
  • ID: oma1al3o
    Snippet: PURPOSE Laparoscopic ventral hernia repair is commonly performed with mesh prostheses; however, there is no standard for fixation devices used to secure mesh to the abdominal wall. This study is a functional comparison of novel, screw-type absorbable and permanent fixation devices with a traditional titanium fixation device. METHODS Fifteen pigs each underwent the laparoscopic placement of two 11 × 14-cm mesh prostheses and were randomized for mesh fixation with either titanium spiral tacks (TS
    Document: PURPOSE Laparoscopic ventral hernia repair is commonly performed with mesh prostheses; however, there is no standard for fixation devices used to secure mesh to the abdominal wall. This study is a functional comparison of novel, screw-type absorbable and permanent fixation devices with a traditional titanium fixation device. METHODS Fifteen pigs each underwent the laparoscopic placement of two 11 × 14-cm mesh prostheses and were randomized for mesh fixation with either titanium spiral tacks (TS), absorbable screw-type fasteners (SF), or permanent screw-type fasteners (PF) (n = 10 mesh prostheses for each fixation group). Adhesions were assessed laparoscopically at 4 weeks. The fixation devices were also embedded in porcine abdominal rectus muscle for ex vivo mechanical testing along with partial thickness polypropylene suture (PR) as a control group (n = 40 for each group). Maximum pull-off forces were measured. All statistical tests were two-tailed, and a P-value < 0.05 was considered to be significant. RESULTS The mean tenacity adhesion scores were 1.40 ± 0.52 (PF), 1.7 ± 0.82 (SF), and 2.6 ± 1.07 (TS). Adhesions in the PF group were significantly less tenacious compared with the TS group (P = 0.01). Quantitative adhesion scores were not significantly different among groups. The maximum pull-off forces, measured in Newtons, were 28.61 N ± 4.89 N (TS), 22.71 N ± 7.86 N (SF), 16.98 N ± 7.59 N (PF), and 20.83 N ± 6.25 N (PR). The pull-off force in the TS group was higher than all of the other groups (P < 0.001). The SF group also had a higher pull-off force compared with the PF group (P < 0.001). CONCLUSIONS The screw-type absorbable and permanent fixation devices provided adequate fixation and were associated with decreased adhesions in this porcine model.

    Search related documents:
    Co phrase search for related documents
    • abdominal wall and adequate fixation provide: 1