Author: Tang, Yi-Wei; Gonsalves, Sarah; Sun, Janet Y; Stiles, Jeffrey; Gilhuley, Kathleen A; Mikhlina, Albina; Dunbar, Sherry A; Babady, N Esther; Zhang, Hongwei
Title: Clinical Evaluation of the Luminex NxTAG Respiratory Pathogen Panel. Cord-id: yer8nr43 Document date: 2016_1_1
ID: yer8nr43
Snippet: An evaluation of the Luminex NxTAG Respiratory Pathogen Panel was performed on 404 clinical respiratory specimens. Clinical sensitivities and specificities of the assay compared to those of the reference methods were 80.0% to 100.0% and 98.9% to 100.0%, respectively. Correct genotyping information was provided for 95.5% of influenza virus A specimens. The closed-tube format of the assay simplified the workflow and minimized carryover contamination.
Document: An evaluation of the Luminex NxTAG Respiratory Pathogen Panel was performed on 404 clinical respiratory specimens. Clinical sensitivities and specificities of the assay compared to those of the reference methods were 80.0% to 100.0% and 98.9% to 100.0%, respectively. Correct genotyping information was provided for 95.5% of influenza virus A specimens. The closed-tube format of the assay simplified the workflow and minimized carryover contamination.
Search related documents:
Co phrase search for related documents- Try single phrases listed below for: 1
Co phrase search for related documents, hyperlinks ordered by date