Author: Engelmann, Ilka; Alidjinou, Enagnon Kazali; Lazrek, Mouna; Dewilde, Anny; Hober, Didier
Title: Necessity to critically review the automatic results of the Xpert Flu assay Cord-id: c813ttt2 Document date: 2017_5_31
ID: c813ttt2
Snippet: Abstract While using the Xpert Flu assay we became aware of false-negative results. The study aimed to analyze the causes of these false-negative results. One hundred fifty-nine respiratory specimens were tested in the Xpert Flu assay and in multiplex reverse transcription–polymerase chain reactions (RT-PCRs) for respiratory viruses. Discordant specimens were tested in the Influenza A/B r-gene assay. One hundred fifty-two (96%) and 151 (95%) specimens yielded concordant results for influenza A
Document: Abstract While using the Xpert Flu assay we became aware of false-negative results. The study aimed to analyze the causes of these false-negative results. One hundred fifty-nine respiratory specimens were tested in the Xpert Flu assay and in multiplex reverse transcription–polymerase chain reactions (RT-PCRs) for respiratory viruses. Discordant specimens were tested in the Influenza A/B r-gene assay. One hundred fifty-two (96%) and 151 (95%) specimens yielded concordant results for influenza A and B, respectively. Fifteen specimens tested negative in the Xpert Flu assay and positive in a multiplex RT-PCR. Positive results were confirmed for 12 of these specimens in the Influenza A/B r-gene assay. Xpert Flu assay amplification curves and endpoints suggested that the false-negative results were mainly due to erroneous automatic result interpretation. We report false-negative results of the Xpert Flu assay due to erroneous automatic result interpretation. Careful analysis of amplification curves and endpoints is needed to avoid reporting of false-negative results.
Search related documents:
Co phrase search for related documents- Try single phrases listed below for: 1
Co phrase search for related documents, hyperlinks ordered by date