Author: Mahesh, Mahadevappa; Siewerdsen, Jeffrey H.
Title: Ultraviolet germicidal irradiation of the inner bore of a CT gantry Cord-id: ez6p4eqg Document date: 2020_11_18
ID: ez6p4eqg
Snippet: PURPOSE: To investigate the feasibility and practicality of ultraviolet (UV) germicidal irradiation of the inner bore of a computed tomography (CT) gantry as a means of viral decontamination. METHOD: A UV lamp (PADNUT 38 W, 253 nm UVâ€C light tube) and UVâ€C dosimeter (GENERAL UVâ€C Digital Light Meter No. UV512C) were used to measure irradiance throughout the inner bore of a CT scanner gantry. Irradiance (units μW/cm(2)) was related to the time required to achieve 6â€log viral kill (10(−
Document: PURPOSE: To investigate the feasibility and practicality of ultraviolet (UV) germicidal irradiation of the inner bore of a computed tomography (CT) gantry as a means of viral decontamination. METHOD: A UV lamp (PADNUT 38 W, 253 nm UVâ€C light tube) and UVâ€C dosimeter (GENERAL UVâ€C Digital Light Meter No. UV512C) were used to measure irradiance throughout the inner bore of a CT scanner gantry. Irradiance (units μW/cm(2)) was related to the time required to achieve 6â€log viral kill (10(−6) survival fraction). RESULTS: A warmâ€up time of ~120 s was required for the lamp to reach stable irradiance. Irradiance at the scan plane (z = 0 cm) of the CT scanner was 580.9 μW/cm(2), reducing to ~350 μW/cm(2) at z = ±20 cm toward the front or back of the gantry. The angular distribution of irradiation was uniform within 10% coefficient of variation. A conservative estimate suggests at least 6â€log kill (survival fraction ≤ 10(−6)) of viral RNA within ±20 cm of the scan plane with an irradiation time of 120 s from cold start. More conservatively, running the lamp for 180 s (3 min) or 300 s (5 min) from cold start is estimated to yield survival fraction <<10(−7) survival fraction within ±20 cm of the scan plane. CONCLUSION: Ultraviolet irradiation of the inner bore of the CT gantry can be achieved with a simple UVâ€C lamp attached to the CT couch. Such practice could augment manual wipeâ€down procedures, improve safety for CT technologists or housekeeping staff, and could potentially reduce turnover time between scanning sessions.
Search related documents:
Co phrase search for related documents- Try single phrases listed below for: 1
Co phrase search for related documents, hyperlinks ordered by date