Author: Bhagwat, M.; Surve, R. M.; Patwardhan, P.
Title: Effect of Heat and Moisture Exchange Filter on the Dead Space Ventilation in a Pediatric Patient Cord-id: uvkwont8 Document date: 2021_1_1
ID: uvkwont8
Snippet: The use of heat and moisture exchange filter (HMEF) has become a routine standard of care in general anesthesia (GA), more so, during the current COVID-19 pandemic times. However its routine use, especially in pediatric patients, has several issues. We report a case of increased dead space ventilation in a pediatric patient due to the presence of HMEF. Hypercapnia and its cause was instantly diagnosed with two capnography sampling lines placed at two ends of the HMEF.
Document: The use of heat and moisture exchange filter (HMEF) has become a routine standard of care in general anesthesia (GA), more so, during the current COVID-19 pandemic times. However its routine use, especially in pediatric patients, has several issues. We report a case of increased dead space ventilation in a pediatric patient due to the presence of HMEF. Hypercapnia and its cause was instantly diagnosed with two capnography sampling lines placed at two ends of the HMEF.