Author: Yang, Wan-hua; Gu, Hong-bin; Chen, Bing; Li, Juan; Fan, Qiu-wei; Yuan, Yong-fang; Wang, Xiangdong
Title: Evaluation of SLOG/TCI-III pediatric system on target control infusion of propofol. Cord-id: bnn3fvm6 Document date: 2011_1_1
ID: bnn3fvm6
Snippet: BACKGROUND The target-controlled infusion-III (SLOG/TCI-III) system was derived from a model set up by the local pediatric population for target control infusion of propofol. METHODS The current study aimed at evaluating the difference between target concentrations of propofol and performance, which was measured using the SLOG/TCI-III system in children. Thirty children fulfilling the I-II criteria according to American Society of Anesthesiology were enrolled in the study. The target plasma conc
Document: BACKGROUND The target-controlled infusion-III (SLOG/TCI-III) system was derived from a model set up by the local pediatric population for target control infusion of propofol. METHODS The current study aimed at evaluating the difference between target concentrations of propofol and performance, which was measured using the SLOG/TCI-III system in children. Thirty children fulfilling the I-II criteria according to American Society of Anesthesiology were enrolled in the study. The target plasma concentration of propofol was fed into the SLOG/TCI-III system and compared with the measured concentrations of propofol. Blood samples were collected and analyzed by high performance liquid chromatography with fluorescence detector. The performance error (PE) was determined for each measured blood propofol concentration. The performances of the TCI-III system were determined by the median performance error (MDPE), the median absolute performance error (MDAPE), and Wobble (the median absolute deviation of each PE from the MDPE), respectively. RESULTS Concentration against target concentration showed good linear correlation: concentration = 1.3428 target concentration - 0.2633 (r = 0.8667). The MDPE and MDAPE of the pediatric system were 10 and 22%, respectively, and the median value for Wobble was 24%. MDPE and MDAPE were less than 15 and 30%, respectively. CONCLUSIONS The performance of TCI-III system seems to be in the accepted limits for clinical practice in children.
Search related documents:
Co phrase search for related documents- Try single phrases listed below for: 1
Co phrase search for related documents, hyperlinks ordered by date