Author: David M Rubin; Neil T Stacey; Tonderayi Matambo; Diane Hildebrandt
Title: Oxygen transfer characteristics of a hollow fiber dialyser: toward possible repurposing of dialysers as blood oxygenators in the context of constrained availability of respiratory support Document date: 2020_4_11
ID: 4igpae7w_30
Snippet: The di culty in applying equation [2] to find kA is twofold: First,the measured oxygen transfer rate represents a minimum due to the limits of our oxygen meter. Secondly, while it is a relatively simple matter to determine an accurate value for the water inlet and air outlet, pO top 2 , it is far more di cult to determine the pO bottom 2 , as we do not have measurements for the oxygen gas pressure as it enters the HFMD......
Document: The di culty in applying equation [2] to find kA is twofold: First,the measured oxygen transfer rate represents a minimum due to the limits of our oxygen meter. Secondly, while it is a relatively simple matter to determine an accurate value for the water inlet and air outlet, pO top 2 , it is far more di cult to determine the pO bottom 2 , as we do not have measurements for the oxygen gas pressure as it enters the HFMD.