Selected article for: "di erence and partial pressure"

Author: David M Rubin; Neil T Stacey; Tonderayi Matambo; Diane Hildebrandt
Title: Oxygen transfer characteristics of a hollow fiber dialyser: toward possible repurposing of dialysers as blood oxygenators in the context of constrained availability of respiratory support
  • Document date: 2020_4_11
  • ID: 4igpae7w_33
    Snippet: Because the outlet water stream is at or above saturation, the partial pressure di erence at that point is at most equal to that over-pressure amount and it may in fact be less. This means that the mass transfer coe cient calculations are highly sensitive to the actual pressure in this region......
    Document: Because the outlet water stream is at or above saturation, the partial pressure di erence at that point is at most equal to that over-pressure amount and it may in fact be less. This means that the mass transfer coe cient calculations are highly sensitive to the actual pressure in this region.

    Search related documents:
    Co phrase search for related documents
    • coe cient and mass transfer: 1, 2, 3, 4
    • coe cient and partial pressure: 1, 2
    • di erence and mass transfer: 1, 2
    • di erence and partial pressure: 1, 2, 3, 4, 5, 6
    • di erence and water stream: 1
    • mass transfer and partial pressure: 1, 2, 3, 4, 5, 6
    • mass transfer and water stream: 1
    • partial pressure and water stream: 1