Selected article for: "CT positive sample and CT value"

Author: Johnson, C.; Ferguson, K.; Smith, T.; Wallace, D.; McConnell, K.; Conserve, D.; martin, i.
Title: EVALUATION OF THE PANBIO SARS-COV-2 RAPID ANTIGEN DETECTION TEST IN THE BAHAMAS
  • Cord-id: mprw4t0y
  • Document date: 2021_7_15
  • ID: mprw4t0y
    Snippet: Identification and isolation of persons infected with SARS-COV-2 are key mitigation strategies in the current pandemic, and rapid antigen detection tests (RADTs) offer the promise of decreased turnaround time to diagnosis when compared with gold standard RT-PCR testing. We evaluated the analytical performance of the Abbott Panbio RADT on nasopharyngeal samples stored at the Ministry of Health National Reference Lab in Nassau, Bahamas. The Panbio demonstrated a test sensitivity of 94% and a speci
    Document: Identification and isolation of persons infected with SARS-COV-2 are key mitigation strategies in the current pandemic, and rapid antigen detection tests (RADTs) offer the promise of decreased turnaround time to diagnosis when compared with gold standard RT-PCR testing. We evaluated the analytical performance of the Abbott Panbio RADT on nasopharyngeal samples stored at the Ministry of Health National Reference Lab in Nassau, Bahamas. The Panbio demonstrated a test sensitivity of 94% and a specificity of 100% on 50 PCR negative samples and 50 samples presumed to be infectious based on having PCR cycle threshold values below 30. Additionally, in our examination of operator results there was low interpersonal variation (1%) among three blind operators and significant correlation between sample Ct value and perceived signal strength on the RADT device. However, three PCR positive samples below Ct 30 were misdiagnosed by RADT, including one sample with Ct value less than 20. These results support the use of the Panbio in symptomatic persons to detect active SARS-COV-2 infections, with the caveat that RADT-negative samples should be confirmed by RT-PCR.

    Search related documents:
    Co phrase search for related documents
    • abbott panbio and active infection: 1
    • abbott panbio and active sars infection: 1
    • abbott panbio and low sensitivity: 1, 2, 3
    • accurately identify and low sensitivity: 1, 2, 3, 4, 5
    • active infection and low sensitivity: 1, 2, 3
    • load sample and low sensitivity: 1, 2, 3, 4