Selected article for: "entrain room air and pressure relief"

Author: Micha Sam Brickman Raredon; Clark Fisher; Paul Heerdt; Ranjit Deshpande; Steven Nivison; Elaine Fajardo; Shamsuddin Akhtar; Thomas Raredon; Laura E. Niklason
Title: Pressure-Regulated Ventilator Splitting (PReVentS): A COVID-19 Response Paradigm from Yale University
  • Document date: 2020_4_6
  • ID: mqj071gp_92
    Snippet: • As the circuit is currently designed, negative circuit pressure near the patient will entrain room air through the negative-pressure relief valve -this will briefly decrease the FiO2 following in-line suctioning, or if the patient makes strong respiratory effort during the ventilator's expiratory phase. If even a brief FiO2 decrease is unacceptable for a patient, an oxygen reservoir bag should be added at the negative pressure relief valve. W.....
    Document: • As the circuit is currently designed, negative circuit pressure near the patient will entrain room air through the negative-pressure relief valve -this will briefly decrease the FiO2 following in-line suctioning, or if the patient makes strong respiratory effort during the ventilator's expiratory phase. If even a brief FiO2 decrease is unacceptable for a patient, an oxygen reservoir bag should be added at the negative pressure relief valve. With proper forethought, steps can also be taken to mitigate FiO2 drop during in-line suctioning (see below). • While not intrinsic to the PReVentS circuit, with the current hand-made valves it is necessary to briefly open the circuit to make valve adjustments. During this maneuver, ventilation is still maintained for the other patient, and PEEP is maintained for both patients. We are currently iterating on valve design to allow pressure adjustment without any circuit disruption.

    Search related documents:
    Co phrase search for related documents
    • circuit open and negative pressure: 1
    • circuit pressure and negative pressure: 1, 2, 3, 4
    • circuit pressure and negative pressure relief: 1, 2, 3, 4
    • circuit pressure and negative pressure relief valve: 1, 2, 3, 4
    • circuit pressure and negative pressure relief valve add: 1
    • circuit pressure and patient maintain: 1
    • currently design and negative pressure: 1
    • line suctioning and negative pressure: 1
    • line suctioning and negative pressure relief: 1
    • line suctioning and negative pressure relief valve: 1
    • line suctioning and oxygen reservoir: 1
    • negative pressure and oxygen reservoir: 1
    • negative pressure relief and oxygen reservoir: 1
    • negative pressure relief valve and oxygen reservoir: 1