Selected article for: "breath stacking and PReVentS circuit"

Author: Micha Sam Brickman Raredon; Clark Fisher; Paul Heerdt; Ranjit Deshpande; Steven Nivison; Elaine Fajardo; Shamsuddin Akhtar; Thomas Raredon; Laura E. Niklason
Title: Pressure-Regulated Ventilator Splitting (PReVentS): A COVID-19 Response Paradigm from Yale University
  • Document date: 2020_4_6
  • ID: mqj071gp_84
    Snippet: As can be seen from careful inspection of Figure 14 , there is some evidence of incomplete exhalation, or "breath stacking", in the mock lung used to simulate Patient #2 in this experiment. Whether this observation is inherent to the PReVentS circuit, or whether it is an artifact of the specific circuit components or the mock lung utilized for this study, is unclear at the present time......
    Document: As can be seen from careful inspection of Figure 14 , there is some evidence of incomplete exhalation, or "breath stacking", in the mock lung used to simulate Patient #2 in this experiment. Whether this observation is inherent to the PReVentS circuit, or whether it is an artifact of the specific circuit components or the mock lung utilized for this study, is unclear at the present time.

    Search related documents: