Author: T. Hans Derr; Melissa A. James; Chad V. Kuny; Prem P. Kandel; Matthew D. Beckman; Kevin L. Hockett; Mark A. Bates; Moriah L Szpara
Title: Aerosolized Hydrogen Peroxide Decontamination of N95 Respirators, with Fit-Testing and Virologic Confirmation of Suitability for Re-Use During the COVID-19 Pandemic Document date: 2020_4_22
ID: d1crwbuz_45
Snippet: Based on a series of ten respirator decontamination cycles thus far, which to date have included two rounds of virologic testing, we found successful respirator decontamination by aHP and fitness for respirator reuse in this pilot study. We found that respirators were not substantively degraded (chemically or physically) by multiple cycles of aHP, as reflected in successful fit testing throughout the decontamination process. We used intentional v.....
Document: Based on a series of ten respirator decontamination cycles thus far, which to date have included two rounds of virologic testing, we found successful respirator decontamination by aHP and fitness for respirator reuse in this pilot study. We found that respirators were not substantively degraded (chemically or physically) by multiple cycles of aHP, as reflected in successful fit testing throughout the decontamination process. We used intentional viral inoculation onto specific respirator surfaces to thoroughly test for the ability of aHP to inactivate viruses, and to discern whether any differences in respirator construction (e.g. outer vs. inner surface fabric)
Search related documents:
Co phrase search for related documents- decontamination process and fit testing: 1, 2, 3, 4
- decontamination process and respirator decontamination: 1, 2, 3, 4, 5, 6, 7
- decontamination process and respirator reuse: 1, 2
- fit testing and pilot study: 1, 2, 3, 4
- fit testing and respirator decontamination: 1, 2, 3, 4, 5, 6, 7, 8
- fit testing and respirator reuse: 1, 2, 3, 4, 5
- fit testing and respirator surface: 1
Co phrase search for related documents, hyperlinks ordered by date