Author: Andrew R Plummer; Jonathan l du Bois; Siu Man Lee; Patrick Magee; Jens Roesner; Harinderjit S Gill
Title: The BathRC model: a method to estimate flow restrictor size for dual ventilation of dissimilar patients Document date: 2020_4_17
ID: cnkldzw9_52
Snippet: A series of tests was carried out to determine the pressure-flow characteristics of the flow restrictors. The restrictors were placed in the inspiration line of a test lung loop so they experienced a realistic unidirectional flow regime, and the two gas flow analysers were inserted in the lines either side of the component (Figure 7) ......
Document: A series of tests was carried out to determine the pressure-flow characteristics of the flow restrictors. The restrictors were placed in the inspiration line of a test lung loop so they experienced a realistic unidirectional flow regime, and the two gas flow analysers were inserted in the lines either side of the component (Figure 7) .