Author: Wan, Yafang; Li, Zhijie; Wang, Kun; Li, Tian; Liao, Pu
Title: Performance verification of anti-SARS-CoV-2-specific antibody detection by using four chemiluminescence immunoassay systems Cord-id: v0971jlq Document date: 2020_11_12
ID: v0971jlq
Snippet: OBJECTIVES: The purpose of the current study was to evaluate the analytical performance of seven kits for detecting IgM/IgG antibodies against coronavirus (SARS-CoV-2) by using four chemiluminescence immunoassay systems. METHODS: Fifty patients diagnosed with SARS-CoV-2 infection and 130 controls without coronavirus infection from the General Hospital of Chongqing were enrolled in the current retrospective study. Four chemiluminescence immunoassay systems, including seven IgM/IgG antibody detect
Document: OBJECTIVES: The purpose of the current study was to evaluate the analytical performance of seven kits for detecting IgM/IgG antibodies against coronavirus (SARS-CoV-2) by using four chemiluminescence immunoassay systems. METHODS: Fifty patients diagnosed with SARS-CoV-2 infection and 130 controls without coronavirus infection from the General Hospital of Chongqing were enrolled in the current retrospective study. Four chemiluminescence immunoassay systems, including seven IgM/IgG antibody detection kits for SARS-CoV-2 (A_IgM, A_IgG, B_IgM, B_IgG, C_IgM, C_IgG and D_Ab), were employed to detect antibody concentrations. The chi-square test, the receiver operating characteristic (ROC) curve and Youden’s index were determined to verify the cut-off value of each detection system. RESULTS: The repeatability verification results of the A, B, C and D systems are all qualified. D_Ab performed best (92% sensitivity and 99.23% specificity), and B_IgM performed worse than the other systems. Except for the A_IgM and C_IgG systems, the optimal diagnostic thresholds and cut-off values of the other kits and their recommendations are inconsistent with each other. B_IgM had the worst AUC, and C_IgG had the best diagnostic accuracy. More importantly, the B_IgG system had the highest false-positive rate for testing patients with AIDS, tumours and pregnancies. The A_IgM system test showed the highest false-positive rates among elderly individuals over 90 years old. COVID-2019 IgM/IgG antibody test systems exhibit performance differences. CONCLUSIONS: The Innodx Biotech Total Antibody serum diagnosis kit is the most reliable detection system for anti-SARS-CoV-2 antibodies, which can be used together with nucleic acid tests as an alternative method for SARS-CoV-2 detecting.
Search related documents:
Co phrase search for related documents- acid result and low specificity: 1
- acid testing and acute period: 1
- acid testing and low sensitivity: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13
- acid testing and low specificity: 1, 2, 3, 4
- acute period and long incubation period: 1, 2
- acute period and low specificity: 1
- long incubation period and low sensitivity: 1
Co phrase search for related documents, hyperlinks ordered by date