Selected article for: "expiratory flow and lung volume"

Author: Uchiyama, Akinori; Yoshida, Takeshi; Yamanaka, Hidenori; Fujino, Yuji
Title: Estimation of tracheal pressure and imposed expiratory work of breathing by the endotracheal tube, heat and moisture exchanger, and ventilator during mechanical ventilation.
  • Cord-id: wt8bldap
  • Document date: 2013_1_1
  • ID: wt8bldap
    Snippet: BACKGROUND The resistance of the endotracheal tube (ETT), the heat and moisture exchanger (HME), and the ventilator may affect the patient's respiratory status. Although previous studies examined the inspiratory work of breathing (WOB), investigation of WOB in the expiratory phase is rare. We estimated tracheal pressure at the tip of the ETT (Ptrach) and calculated expiratory WOB imposed by the ETT, the HME, and the expiratory valve. We examined imposed expiratory WOB in patients under a continu
    Document: BACKGROUND The resistance of the endotracheal tube (ETT), the heat and moisture exchanger (HME), and the ventilator may affect the patient's respiratory status. Although previous studies examined the inspiratory work of breathing (WOB), investigation of WOB in the expiratory phase is rare. We estimated tracheal pressure at the tip of the ETT (Ptrach) and calculated expiratory WOB imposed by the ETT, the HME, and the expiratory valve. We examined imposed expiratory WOB in patients under a continuous mandatory ventilation (CMV) mode and during spontaneous breathing trials (SBTs). We hypothesized that imposed expiratory WOB would increase with heightened ventilatory demand. METHODS We measured airway pressure (Paw) and respiratory flow (V). We estimated Ptrach using the equation Ptrach = Paw - K1 × V(K2) - 2.70 × V(L/s)(1.42). K1 and K2 were determined by the inner diameter (ID) of the ETT. Imposed expiratory WOB was calculated from the area of Ptrach above PEEP versus lung volume. We examined imposed expiratory WOB and imposed expiratory resistance in relation to mean expiratory flow. RESULTS We examined 28 patients under CMV mode, and 29 during SBT. During both CMV and SBT, as mean expiratory flow increased, imposed expiratory WOB increased. The regression curves between mean expiratory flow (x) (L/s) and imposed expiratory WOB (y) (J/L) were y = 1.35x(0.83) (R(2) = 0.79) for 7 mm ID ETT under CMV, y = 1.12x(0.82) (R(2) = 0.73) for 8 mm ID ETT under CMV, y = 1.07x(1.04) (R(2) = 0.85) for 7 mm ID ETT during SBT, and y = 0.84x(0.93) (R(2) = 0.75) for 8 mm ID ETT during SBT. Levels of imposed expiratory WOB were affected by ETT diameter and ventilator mode. The reason for increasing imposed expiratory WOB was an increase in expiratory resistance imposed by the ETT and HME. CONCLUSIONS Under mechanical ventilation, imposed expiratory WOB should be considered in patients with higher minute ventilation.

    Search related documents:
    Co phrase search for related documents
    • Try single phrases listed below for: 1
    Co phrase search for related documents, hyperlinks ordered by date