Selected article for: "experiment goal and inspiratory pressure"

Author: Micha Sam Brickman Raredon; Clark Fisher; Paul Heerdt; Ranjit Deshpande; Steven Nivison; Elaine Fajardo; Shamsuddin Akhtar; Thomas Raredon; Laura E. Niklason
Title: Pressure-Regulated Ventilator Splitting (PReVentS): A COVID-19 Response Paradigm from Yale University
  • Document date: 2020_4_6
  • ID: mqj071gp_74
    Snippet: For this experiment, the circuit was connected to two mock lungs having similar compliance. The goal of the experiment was to tune settings of the inspiratory pressure-gated valve for one lung, and to measure the resultant tidal volumes for both of the lungs in the PReVentS circuit ( Figure 12 ). As is clear from Figure 12 , changing the pressure setting on one valve in the inspiratory limb for Patient #1 had a linear effect on the resultant tida.....
    Document: For this experiment, the circuit was connected to two mock lungs having similar compliance. The goal of the experiment was to tune settings of the inspiratory pressure-gated valve for one lung, and to measure the resultant tidal volumes for both of the lungs in the PReVentS circuit ( Figure 12 ). As is clear from Figure 12 , changing the pressure setting on one valve in the inspiratory limb for Patient #1 had a linear effect on the resultant tidal volume of Patient #1, while it had minimal to no effect on the measured tidal volume for Patient #2. The PReVentS . CC-BY-NC-ND 4.0 International license It is made available under a author/funder, who has granted medRxiv a license to display the preprint in perpetuity.

    Search related documents:
    Co phrase search for related documents
    • cc NC ND International license and inspiratory limb: 1, 2
    • cc NC ND International license and International license: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
    • cc NC ND International license and Patient tidal volume: 1, 2, 3
    • cc NC ND International license and PReVentS circuit: 1, 2, 3
    • cc NC ND International license and similar compliance: 1, 2
    • cc NC ND International license and tidal volume: 1, 2, 3, 4, 5, 6
    • experiment goal and inspiratory limb: 1, 2
    • experiment goal and International license: 1, 2
    • experiment goal and Patient tidal volume: 1
    • experiment goal and similar compliance: 1, 2, 3, 4
    • experiment goal and tidal volume: 1, 2
    • inspiratory limb and International license: 1, 2
    • inspiratory limb and mock lung: 1
    • inspiratory limb and Patient tidal volume: 1, 2, 3, 4, 5
    • inspiratory limb and PReVentS circuit: 1
    • inspiratory limb and similar compliance: 1, 2
    • inspiratory limb and tidal volume: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12
    • International license and linear effect: 1, 2
    • International license and Patient tidal volume: 1, 2, 3, 4, 5