Author: Hongquan Li; Ethan Li; Deepak Krishnamurthy; Patrick Kolbay; Beca Chacin; Soeren Hoehne; Jim Cybulski; Lara Brewer; Tomasz Petelenz; Joseph Orr; Derek Sakata; Thomas Clardy; Kai Kuck; Manu Prakash
Title: Utah-Stanford Ventilator (Vent4US): Developing a rapidly scalable ventilator for COVID-19 patients with ARDS Document date: 2020_4_22
ID: cfpd45qf_7
Snippet: respiratory rate, as well as assist control where a breath of set tidal volume can be triggered by the patient. The linear actuator ( Figure 4A ) compresses a segment of 1/4" OD, 1/8" ID silicone tubing (hardness: Shore A45) by variable amounts to achieve flow control. At an inlet pressure of 15 psi, we found that a 3 mm actuation is sufficient for valve flow to transition from a completely closed state to a completely open state ( Figure 4B and .....
Document: respiratory rate, as well as assist control where a breath of set tidal volume can be triggered by the patient. The linear actuator ( Figure 4A ) compresses a segment of 1/4" OD, 1/8" ID silicone tubing (hardness: Shore A45) by variable amounts to achieve flow control. At an inlet pressure of 15 psi, we found that a 3 mm actuation is sufficient for valve flow to transition from a completely closed state to a completely open state ( Figure 4B and 4C), with the completely open state allowing the maximum flow rate of 50 L/min. We also demonstrated both slow and fast actuation cycles of the valve ( Figure 4D and 4E, at actuation rates for fully-closed to fully-open cycles up to 5 Hz). We measured the valve's response time (time it takes for flow to reach 95% of its maximum value) to be 48˘8 ms ( Figure 4F ). Finally we tested the repeatability of the valve's proportional flow control by measuring the flow rate as a function of valve state ( Figure 4G ). The valve was found to be fairly repeatable, with the flow rate at a given valve state having a maximum variability of around 5% at the highest flow rates. Repeatability can be further improved by incorporating an encoder or a Hall effect limit switch.
Search related documents:
Co phrase search for related documents- maximum value and respiratory rate: 1, 2, 3
- maximum value and tidal volume: 1
- patient trigger and response time: 1
- proportional flow control and valve flow: 1
- proportional flow control and valve proportional flow control: 1
- respiratory rate and tidal volume: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 57, 58, 59, 60, 61, 62, 63, 64, 65, 66, 67, 68, 69, 70, 71, 72, 73
- respiratory rate and valve flow: 1, 2
- response time and tidal volume: 1
- silicone tube and tidal volume: 1
- tidal volume and valve flow: 1, 2, 3, 4, 5, 6, 7, 8
Co phrase search for related documents, hyperlinks ordered by date