Selected article for: "clinical sample and PCR assay"

Author: Graham, Maryza; Williams, Eloise; Isles, Nicole; Buadromo, Eka; Toatu, Tebuka; Druce, Julian; Catton, Mike; Lin, Chantel; Howden, Benjamin P.; Williamson, Deborah A.
Title: Sample pooling on the Cepheid Xpert® Xpress SARS-CoV-2 assay
  • Cord-id: cqze6607
  • Document date: 2020_10_15
  • ID: cqze6607
    Snippet: The COVID-19 pandemic has placed unprecedented global demand on laboratory supplies required for testing. Sample pooling has been investigated by laboratories as a strategy to preserve testing capacity. We evaluate the performance of Cepheid Xpert® Xpress SARS-CoV-2 RT-PCR assay for testing samples in pools of four and six. Clinical samples containing SARS-CoV-2, and confirmed negative clinical samples were used to create sample pools. Clinical samples had ‘neat’ Xpert® E gene cycle thresh
    Document: The COVID-19 pandemic has placed unprecedented global demand on laboratory supplies required for testing. Sample pooling has been investigated by laboratories as a strategy to preserve testing capacity. We evaluate the performance of Cepheid Xpert® Xpress SARS-CoV-2 RT-PCR assay for testing samples in pools of four and six. Clinical samples containing SARS-CoV-2, and confirmed negative clinical samples were used to create sample pools. Clinical samples had ‘neat’ Xpert® E gene cycle threshold (Ct) values ranging between 20 and 28 and all were detected qualitatively when contained in pools of four or six samples. For these samples, pooling had a median change in Ct value of 2.0 in pools of four, and of 2.9 in pools of six. With the use of Cepheid Xpert® Xpress SARS-CoV-2 RT-PCR assay, pooling of four or six samples may be an effective strategy to increase testing capacity.

    Search related documents:
    Co phrase search for related documents
    • accurate rapid and low dilution: 1
    • accurate rapid and low resource: 1, 2, 3, 4, 5, 6
    • accurate rapid and low sample: 1, 2
    • lod detection and low resource: 1, 2, 3
    • lod detection and low sample: 1, 2, 3
    • lod detection and low sample input: 1