Title: Ad-hoc Assembly of Lean Extracorporeal Membrane Oxygenation Systems for COVID-19
  • Document date: 2020_4_17
  • ID: 41v5shpt_11
    Snippet: In the dual gas exchanger setup that uses a first gas exchange module in-line to partially deoxygenate the blood and then uses the second gas exchanger to re-oxygenate the blood, we observed a stable, near-maximal oxygen content after the second oxygenator over the observation time of 2 days, although the experiments performed at room temperatures and with N2 for O2 washout may not fully reflect physiological conditions......
    Document: In the dual gas exchanger setup that uses a first gas exchange module in-line to partially deoxygenate the blood and then uses the second gas exchanger to re-oxygenate the blood, we observed a stable, near-maximal oxygen content after the second oxygenator over the observation time of 2 days, although the experiments performed at room temperatures and with N2 for O2 washout may not fully reflect physiological conditions.