Author: Patrick Hunziker; Urs Zenklusen
Title: Ad-hoc Assembly of Lean Extracorporeal Membrane Oxygenation Systems for COVID-19 Document date: 2020_4_17
ID: 41v5shpt_11
Snippet: In the dual gas exchanger setup that uses a first gas exchange module in-line to partially deoxygenate the blood and then uses the second gas exchanger to re-oxygenate the blood, we observed a stable, near-maximal oxygen content after the second oxygenator over the observation time of 2 days, although the experiments performed at room temperatures and with N2 for O2 washout may not fully reflect physiological conditions......
Document: In the dual gas exchanger setup that uses a first gas exchange module in-line to partially deoxygenate the blood and then uses the second gas exchanger to re-oxygenate the blood, we observed a stable, near-maximal oxygen content after the second oxygenator over the observation time of 2 days, although the experiments performed at room temperatures and with N2 for O2 washout may not fully reflect physiological conditions.
Search related documents:
Co phrase search for related documents- gas exchange and observation time: 1, 2
- gas exchange and oxygen content: 1, 2, 3
- gas exchange and physiological condition: 1
- gas exchange module and observation time: 1
Co phrase search for related documents, hyperlinks ordered by date