Selected article for: "copy number and human gdna"

Author: Andrew C Nelson; Benjamin Auch; Matthew Schomaker; Daryl M Gohl; Patrick Grady; Darrell Johnson; Robyn Kincaid; Kylene E Karnuth; Jerry Daniel; Jessica K Fiege; Elizabeth J Fay; Tyler Bold; Ryan A Langlois; Kenneth B Beckman; Sophia Yohe
Title: Analytical Validation of a COVID-19 qRT-PCR Detection Assay Using a 384-well Format and Three Extraction Methods
  • Document date: 2020_4_5
  • ID: ec3egn8o_26
    Snippet: We utilized the synthetic positive extraction control and a synthetic negative extraction control for validation of the lower limit of detection for the assay. These extraction controls are delivered in a synthetic transport matrix including human gDNA. We diluted 200 viral cp/µL standard control into the negative extraction control matrix containing 75 human gDNA copies/µL to achieve final viral targets of 20, 10, and 5 copies per microliter. .....
    Document: We utilized the synthetic positive extraction control and a synthetic negative extraction control for validation of the lower limit of detection for the assay. These extraction controls are delivered in a synthetic transport matrix including human gDNA. We diluted 200 viral cp/µL standard control into the negative extraction control matrix containing 75 human gDNA copies/µL to achieve final viral targets of 20, 10, and 5 copies per microliter. We performed 2 to 4 extractions of aliquots at each copy number target using a combination of Qiagen and Macherey Nagel extraction kits; 2 to 5 PCR replicates were performed for each unique extraction. At 5 viral copies/µL input to extraction, we successfully detected N1 and N2 in 20 of 20 replicates (Table 8) , establishing a reliable limit of detection (LOD). The average N1 Ct at this LOD was 36.0 for the Qiagen extraction and 36.7 for the Macherey Nagel extraction; the average N2 Ct values were 35.8 and 37.2 respectively. Overall, the experiment had 97% accuracy at 1X to 2X of the limit of detection (5-10 cp/µL). One of ten PCR replicates at the 10 cp/µL input level returned an inconclusive result, detecting N1 but not N2.

    Search related documents:
    Co phrase search for related documents
    • assay detection low limit and detection low limit: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10
    • assay detection low limit and LOD detection: 1, 2
    • assay detection low limit and low limit: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10
    • copy number and Ct value: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11
    • copy number and detection limit: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20
    • copy number and extraction control: 1
    • copy number and extraction kit: 1, 2, 3, 4, 5
    • copy number and human gdna: 1
    • copy number and LOD detection: 1, 2, 3, 4, 5, 6
    • copy number target and detection limit: 1, 2
    • Ct value and detection limit: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
    • Ct value and extraction control: 1, 2, 3, 4, 5, 6, 7
    • Ct value and extraction kit: 1, 2, 3, 4, 5, 6, 7
    • Ct value and inconclusive result: 1, 2
    • Ct value and LOD detection: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10
    • detection limit and inconclusive result: 1, 2
    • detection limit and input level: 1
    • detection limit and LOD detection: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25
    • detection limit and low limit: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25