Author: Kim, Mi-Na; Ko, Young Jin; Seong, Moon-Woo; Kim, Jae-Seok; Shin, Bo-Moon; Sung, Heungsup
Title: Analytical and Clinical Validation of Six Commercial Middle East Respiratory Syndrome Coronavirus RNA Detection Kits Based on Real-Time Reverse-Transcription PCR Document date: 2016_6_24
ID: 1le537yd_9
Snippet: The effect of PCR inhibition determined by testing the nine high inhibition specimens revealed that 3-5 of the six known positive specimens were found negative or invalid by each kit ( Table 3 ). The Ct values of the internal controls from all specimens were high or extended beyond the total amplification cycles, and, therefore, false negative results were eliminated owing to internal control failure, except for those of the PowerChek and AccuPow.....
Document: The effect of PCR inhibition determined by testing the nine high inhibition specimens revealed that 3-5 of the six known positive specimens were found negative or invalid by each kit ( Table 3 ). The Ct values of the internal controls from all specimens were high or extended beyond the total amplification cycles, and, therefore, false negative results were eliminated owing to internal control failure, except for those of the PowerChek and AccuPower kits. With the AccuPower kit, the mean internal control Ct value for the high inhibition specimens was considerably prolonged to 30.1 ± 0.69 compared with 25.7 ± 0.31 for the sensitivity test specimens; however, there was no established cut-off point, indicating that PCR inhibition was not acceptable in the PCR reaction. On the other hand, Ct values for the sensitivity test specimens and the high inhibition specimens using the PowerChek kit were not significantly different at 25.3 ± 3.2 and 27.0 ± 3.1, respectively. The LightMix kit produced two uncertain results with known positive specimens and one uncertain result with a known negative specimen (Table 3) . There were no positive results detected by any of the kits among the four negative specimens included in the sensitivity test and high inhibition panels (Tables 2 and 3) .
Search related documents:
Co phrase search for related documents- amplification cycle and Ct value: 1, 2, 3, 4, 5
- Ct value and cut point: 1, 2
- Ct value and false negative result: 1, 2, 3, 4
- Ct value and high inhibition: 1
- Ct value and internal control: 1, 2, 3, 4, 5, 6, 7, 8, 9
- cut point and establish cut point: 1, 2, 3, 4, 5
- false negative result and internal control: 1, 2
- high inhibition and inhibition panel: 1, 2
- high inhibition and internal control: 1, 2
- high inhibition panel and inhibition panel: 1, 2
- high inhibition panel sensitivity test and inhibition panel: 1
Co phrase search for related documents, hyperlinks ordered by date