Author: Kim, Mi-Na; Ko, Young Jin; Seong, Moon-Woo; Kim, Jae-Seok; Shin, Bo-Moon; Sung, Heungsup
Title: Analytical and Clinical Validation of Six Commercial Middle East Respiratory Syndrome Coronavirus RNA Detection Kits Based on Real-Time Reverse-Transcription PCR Document date: 2016_6_24
ID: 1le537yd_7
Snippet: The LODs for upE varied from 21.88 to 263.03 copies/reaction, and those of ORF1a varied from 6.92 to 128.82 copies/reaction (Fig. 1) . According to the probit regression analysis, the 95% confidence intervals (CI) for upE and ORF1a were found to overlap among the tested kits, with the exception of ORF1a by the PowerChek kit (Fig. 2) . The LODs for upE using both the single and multiple gene-targeting formats of the 2-step kits were 1.64 and 1.45 .....
Document: The LODs for upE varied from 21.88 to 263.03 copies/reaction, and those of ORF1a varied from 6.92 to 128.82 copies/reaction (Fig. 1) . According to the probit regression analysis, the 95% confidence intervals (CI) for upE and ORF1a were found to overlap among the tested kits, with the exception of ORF1a by the PowerChek kit (Fig. 2) . The LODs for upE using both the single and multiple gene-targeting formats of the 2-step kits were 1.64 and 1.45 log copies/reaction for the PowerChek kit and 1.76 and 1.61 log copies/reaction for the DiaPlexQ kit (Fig. 2 ). The LOD for upE was 1.34 log copies/reaction for the Any-plex kit, but no CI value could be calculated because there were no positive reactions at 1.0 log copy/reaction, while all 16 replicate specimens were positive at 1.5 log copies/reaction. In contrast, although the LODs for upE and ORF1a using the LightMix kit were 2.11 log copies/reaction and 1.78 log copies/reaction, respectively, trailing of positives was observed at much lower concentrations than these LODs (Fig. 1) . The LODs for upE of three different 1-step kits, the AccuPower, Light Mix, and Ultrafast kits, were > 2.0 log copies/reaction, which was less sensitive than those of three different 2-step kits; however, this difference was not significant (Fig. 2) . The LODs for ORF1a varied less-i.e., from 1.78 to 2.11 log copies/reaction-with the exception of the PowerChek kit, which was considerably lower at 0.84 log copies/reaction (Fig. 2) . None of the kits tested in this study showed cross-reactivity with other respiratory viruses.
Search related documents:
Co phrase search for related documents- low concentration and study test: 1, 2, 3, 4, 5
- low concentration and test kit: 1, 2
- ORF1a exception and PowerChek kit: 1
- ORF1a exception and PowerChek kit ORF1a exception: 1
- ORF1a exception and PowerChek kit reaction: 1
- ORF1a upE lod and probit regression: 1
- ORF1a upE lod and probit regression analysis: 1
- ORF1a upE lod and respiratory virus: 1
- ORF1a upE lod and upE lod: 1, 2
- plex kit and respiratory virus: 1
- plex kit reaction and respiratory virus: 1
- positive reaction and respiratory virus: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 57, 58, 59, 60, 61, 62, 63, 64, 65, 66, 67, 68, 69, 70, 71
- positive reaction and study test: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19
- positive reaction and test kit: 1, 2
- PowerChek kit and respiratory virus: 1
- PowerChek kit and test kit: 1
Co phrase search for related documents, hyperlinks ordered by date